A solution provider focusing on infant oxygen inhalation and breathing equipment
contact us
One-touch dial
One-click navigation
Wechat
Rirui Medical
logo
A solution provider focusing on infant oxygen inhalation and breathing equipment

Home > Product > mixed instrument

RL-6000B

release time: 2019-05-27 16:36:57

Views: 278


Registration certificate number: jiangxi machinery note 20162540325


Products overview:

This equipment is based on noninvasive continuous positive airway pressure (NCPAP), designed specifically for pediatric respiratory support medical equipment, in pediatrics for premature infants, newborns, infants and children under the age of 14, the implementation of nasal mask ventilation (NCPAP) when oxygen concentration monitoring, regulation and function of wetting. Equipment flow and the oxygen concentration regulation control the proportion of electronic empty oxygen mix technology, to ensure the accuracy of the output value maximization.


Alarm function: gas source input fault, electricity shortage, high airway pressure, airway pressure is low, the high oxygen concentration alarm, low oxygen concentration alarm;


Alarm mute: no more than 120 s;


The maximum limit pressure: 60 cmH2O or less;


Performance characteristics:

Using liquid crystal display with touch function, the operating interface is concise and straightforward;

Stable oxygen concentration adjustable gas supply, constant flow design;

CPAP pressure can be stable, precise adjustment;

Inhaled air and humid, and the gas constant temperature automatic control;

Real-time monitor the gas oxygen density, flow value;

A variety of alarm functions, to ensure safe and reliable ventilation;

A built-in battery, ensure continuous power supply.


Technical parameters:

Inhaled oxygen concentration regulation and monitoring scope: 21% to 100%.

Flow regulating and monitoring the range: 1 L/min - 20 L/min;

Airway pressure regulating range: 2 cmH2O - 20 cmH2O

CmH2O airway pressure monitoring scope: 0-60 cmH2O

High oxygen concentration to oxygen: supplies with 100% oxygen concentration, maintain for 2 min.